Frequency range |
| 2MHzto3GHz |
|
with R&S®ZPH-B4 option installed | 2 MHz to 4GHz |
|
Frequency resolution |
| 1Hz |
|
Reference frequency, internal |
|
Total reference accuracy |
| ±(time since last adjustment·aging rate) + temperature drift + calibration accuracy |
|
Aging per year |
| ±1·10–6 |
|
Temperature drift | 0℃ to +30 ℃ | ±1·10–6 |
|
+30℃ to +50℃ | ±3·10–6 |
|
Achievable initial calibration accuracy |
| ±5·10–7 |
|
Individual |
| reflection(S11) |
|
| one-port cable loss |
|
| distance-to-fault |
|
model.12 | transmission(S21) |
|
Measurement wizard |
|
Guides the user through a sequence of individual measurements.Uses the R&S®Instrument View PC software to configure the measurement sequence including hints displayed on the screen. R&S®Instrument View is also used to combine the measurement results into user-configurable reports. |
|
Measurement setup |
|
Port output power | model.02 | –10dBm(nom.) |
|
model.12 | –3dBmto –30dBm step 1dB(nom.) |
|
Data points | selectable | 101to2501 |
|
Measurement bandwidth | reflection measurement(S11) | 10kHz |
|
transmission measurement(S21) | 10Hz,100Hz,1kHz,10kHz,100kHz, 1MHz |
|
Trace modes |
| clear/write,average |
|
Reflection measurement S11 |
|
Result formats |
| magnitude, SWR, magnitude and distance-to-fault, SWR and distance-to-fault, smith chart, phase |
|
Magnitude |
|
Range |
| 1/2/3/5/10/20/30/50/100/120/130/150 dB, linear 100% |
|
Resolution |
| 0.1dB |
|
SWR |
|
Range | selectable | 1 to 1.1/1.5/2/3/6/11/21/71 |
|
Measurement speed |
| 0.3msperpoint |
|
Corrected directivity with R&S®ZN-Z103 | 2MHz≤f≤4GHz | >42dB(nom.) |
|
Corrected test port match with R&S®ZN-Z103 | 2MHz≤f≤4GHz | >36dB(nom.) |
|
Measurement uncertainty with R&S®ZN-Z103 |
| see figure Uncertainty of reflection measurement |
|
One-port cable loss measurement |
|
Result format |
| magnitude |
|
Range | selectable | 1/2/5/10/20/50/100dB |
|
Resolution |
| 0.1dB |
|
Distance-to-fault analysis |
|
Result formats |
| return loss, SWR, split screen DTF and SWR, split screen DTF and return loss |
|
Return loss |
|
Range |
| 1/2/3/5/10/20/30/50/100/120/130/150 dB,linear 100% |
|
Resolution |
| 0.01dB |
|
SWR |
|
Range | selectable | 1to1.1/1.5/2/3/6/11/21/71 |
|
Fault resolution |
| (1.5·108·velocity factor/span)m |
|
Maximum cable length | depending on cable loss | 1500m(nom.) |
|
Immunity to interference |
|
Maximum permissible spurious signal | measurement =reflection(S11)/one-port cable loss/distance-to-fault analysis |
|
| +17dBm(nom.) |
|
Accuracy of reflection measurements |
|
2MHzto4GHz | 0dBto–15dB | <0.3dB or <2.1 |
|
–15dBto–25dB | <1.0dB or <7.5 |
|
–25dBto–35dB | <3.1dB or <26 |
|
Dynamic of transmission measurements (model .12 only) |
|
RF Attenuation =5dB,trackinggenerator level=–3dBm,RBW=10kHz | 100kHz≤f<20MHz | >60dB(nom.) |
|
20MHz≤f<1.5GHz | >90dB(nom.) |
|
1.5GHz≤f<2.5GHz | >70dB(nom.) |
|
2.5GHz≤f<4GHz | >60dB(nom.) |
|
Frequency range |
| 5kHz to 3GHz |
|
with R&S®ZPH-B4 option installed | 5kHz to 4GHz |
|
Frequency resolution |
| 1Hz |
|
Reference frequency, internal |
|
Aging per year |
| 1·10–6 |
|
Temperature drift | 0℃to+50℃ | 1·10–6 |
|
Achievable initial calibration accuracy |
| 5·10–7 |
|
Total reference uncertainty |
| (time since lastadjustment·agingrate)+temperature drift+calibration accuracy |
|
Frequency readout |
|
Marker resolution |
| 1Hz |
|
Uncertainty |
| ±(marker frequency·referenceuncertainty+10%·resolutionbandwidth +½(span/(sweeppoints–1)+1Hz) |
|
Number of sweep (trace) points |
| 711 |
|
Marker tuning frequency step size |
| span/710 |
|
Frequency counter resolution |
| 0.1Hz |
|
Count uncertainty | SNR>25dB | ±(frequency·reference uncertainty+ ½(lastdigit)) |
|
Frequency span |
| 0Hz,10Hzto3GHz |
|
with R&S®ZPH-B4 option installed | 0Hz,10Hzto4GHz |
|
Span uncertainty |
| 1%(nom.) |
|
Spectral purity SSB phase noise |
| f=500MHz |
|
Carrier offset | 30kHz | <–88dBc(1Hz),–95dBc(1Hz)(typ.) |
|
100kHz | <–98dBc(1Hz),–105dBc(1Hz)(typ.) |
|
1MHz | <–118dBc(1Hz),–125dBc(1Hz)(typ.) |
|
Sweep time | span=0Hz | 1msto1000s |
|
10Hz≤span≤600MHz | 20ms to 1000s |
|
span>600MHz | 20ms·span/1600MHz to 1000s |
|
Uncertainty | span=0Hz | 1%(nom.) |
|
span≥10Hz | 3%(nom.) |
|
Resolution bandwidths |
|
Range | –3dB bandwidths | 1Hz to 3MHzin1/3sequence |
|
Bandwidth accuracy | 1Hz≤RBW≤300kHz | <5%(nom.) |
|
300kHz<RBW≤1MHz | <10%(nom.) |
|
Selectivity 60dB:3dB |
| <5(nom.),Gaussian type filters |
|
Video filters |
|
Range | –3dBbandwidths | 1Hz to 3MHz in1/3sequence |
|
Display range |
| displayed noise floor to +30dBm |
|
Maximum rated input level |
|
DC voltage |
| 50V |
|
CW RF power |
| 33dBm(=2W) |
|
Peak RF power | duration<3s | 36dBm(=4W) |
|
Maximum pulse voltage |
| 150V |
|
Maximum pulse energy | pulse width10µs | 10mWs |
|
Intermodulation |
|
Third-order intercept (TOI) | intermodulation-free dynamic range, signal level–20dBm(both), RF attenuation=0dB,RFpreamplifier=off |
|
f=1GHz | +7dBm(meas.) |
|
f=2.4GHz | +10dBm(meas.) |
|
Second-harmonic intercept (SHI) | RF attenuation = 0 dB, RF preamplifier = off, signal level = –40 dBm |
|
fin= 20 MHz to 1.5 GHz | –60 dBc (nom.) |
|
fin= 1.5 GHz to 2 GHz | –80 dBc (nom.) |
|
Displayed average noise level | 0 dB RF attenuation, termination 50 Ω, RBW = 1 kHz, VBW = 10 Hz, sample detector, logarithmic scaling, normalized to 1 Hz |
|
preamplifier = off |
|
1 MHz to 10 MHz | < –130 dBm, –135 dBm (typ.) |
|
10 MHz to 1 GHz | < –142 dBm, –146 dBm (typ.) |
|
1 GHz to 4 GHz | < –140 dBm, –144 dBm (typ.) |
|
preamplifier = on |
|
1 MHz to 10 MHz | < –150 dBm, –160 dBm (typ.) |
|
10 MHz to 3 GHz | < –158 dBm, –163 dBm (typ.) |
|
3 GHz to 4 GHz | < –156 dBm, –161 dBm (typ.) |
|
Immunity to interference, nominal values |
|
Image frequencies | fin– 2 · 30.15 MHz | < –70 dBc (nom.) |
|
fin– 2 · 830.15 MHz | < –70 dBc (nom.) |
|
f < 3 GHz, fin– 2 · 830.15 MHz | < –70 dBc (nom.) |
|
f < 3 GHz, fin– 2 · 4042.65 MHz | –60 dBc (nom.) |
|
f ≥ 3 GHz, fin+ 2 · 830.15 MHz | –60 dBc (nom.) |
|
Intermediate frequencies | 30.15 MHz, 830.15 MHz, 4042.65 MHz | < –60 dBc (nom.) |
|
Other interfering signals, signal level – RF attenuation < –30 dBm | f ≤ 3 GHz, spurious at fin– 2021.325 MHz | < –60 dBc (nom.) |
|
Other interfering signals, related to local oscillators | ∆f ≥ 300 kHz | < –60 dBc (nom.) |
|
f = receive frequency |
|
Residual spurious response | input matched with 50 Ω, without input signal, RBW ≤ 30 kHz, f ≥ 3 MHz, RF attenuation = 0 dB | < –90 dBm (nom.) |
|
Level display |
|
Logarithmic level axis |
| 1/2/3/5/10/20/30/50/100/120/150 dB, 10 divisions |
|
Linear level axis |
| 0 % to 100 %, 10 divisions |
|
Number of traces |
| 2 |
|
Trace detectors |
| max. peak, min. peak, auto peak, sample, RMS |
|
Trace functions |
| clear/write, max. hold, min. hold, average, view |
|
Setting range of reference level |
| –130 dBm to +30 dBm |
|
Units of level axis |
| dBm, dBmV, dBµV, V, W |
|
Level measurement uncertainty |
|
Absolute level uncertainty at 100 MHz | +20 ℃ to +30 ℃ | < 0.3 dB |
|
Frequency response (+20 ℃ to +30 ℃) | 5 kHz ≤ f < 10 MHz | < 1.5 dB (nom.) |
|
10 MHz ≤ f ≤ 4 GHz | < 1 dB |
|
Attenuator uncertainty |
| < 0.3 dB |
|
Uncertainty of reference level setting |
| < 0.1 dB (nom.) |
|
Display nonlinearity | SNR > 16 dB, 0 dB to –50 dB, logarithmic level display | < 0.3 dB |
|
Bandwidth switching uncertainty | reference: RBW = 10 kHz | < 0.1 dB (nom.) |
|
Total measurement uncertainty | 95 % confidence level, +20 ℃ to +30 ℃, SNR > 16 dB, 0 dB to –50 dB below reference level, RF attenuation auto |
|
10 MHz ≤ f ≤ 4 GHz | < 1.25 dB, 0.5 dB (typ.) |
|
Trigger |
Trigger source |
| free run, video, external |
External trigger level threshold | low → high transition | 2.4 V |
high → low transition | 0.7 V |
maximum | 3.0 V |
Gated trigger (model .12 only) |
Gate delay |
| 1 µs to 100 s, resolution min. 1 µs (or 1 % of delay) |
Gate length |
| 1 µs to 100 s, resolution min. 1 µs (or 1 % of gate length) |